Product Details:
Payment & Shipping Terms:
|
Mass Median Diameter: | 0.1μm | Quantity Median Diameter: | 0.20μm |
---|---|---|---|
Aerosol Concentration: | 50 ~ 200mg / M3 | Detection Method: | Dual Solid-state Photometer |
Flow Range: | 15 ~ 100 L / Min |
ASTM F2100 Standard Mask testing, Particle Filtration Efficiency Tester (PFE)
Standard Reference
1. YY0469-2011 "Medical surgical mask"
2. ASTM F2100
3. GB2626-2006 "Respiratory protective equipment self-priming filter anti-particulate respirator"
4. EN149-2001+A1-2001 “Respiratory protective devices - Filtering half masks to protect against particles - Requirements, testing, marking”
5. "European Standard EN1822-3: 2009 Single Sheet Filter Test"
The test bench uses a polydisperse experimental aerosol (DEHS or NaCl) to quantify the dust. Under the action of a clean air flow (negative pressure), the conventional performance test is carried out through the sampling of aerosol particles and the pressure difference measurement point of the filter media under test .
The test bench refers to the advanced design concepts of similar international equipment (such as the 8130 model of the US TSI company), but it is also suitable for the test status of related domestic industries. The independently developed program software runs stably and reliably and is easy to operate. Simple operation interface, the operation parameters are clear at a glance. Different dust sources (DEHS and NaCl) can be switched freely. It is suitable for testing a variety of air filter materials
(meltblown, glass fiber, PP, film and other filter materials), medical masks and some labor protection products.
Main technical parameters:
1.Aerosol occurrence:
Oily type | Salt type | |
Aerosol type | PAO, DOP, DEHS, paraffin | sodium chloride |
Mass median diameter | 0.1μm | 0.26μm |
Quantity median diameter | 0.20μm | 0.075μm |
Geometric standard deviation | <1.6 | <1.83 |
Aerosol concentration | 50 ~ 200mg / m3 | 12 ~ 20mg / m3 |
2.Aerosol detection
Detection method | dual solid-state photometer |
Concentration measurement range | 1.0μ g / m3 ~ 200mg / m3 |
Flow range | 15 ~ 100 L / min |
Efficiency measurement range | 0 ~ 99.999% |
Requirements for compressed air source | flow rate 198 L / min at 550 kPa |